Opin vísindi

Automation improves repeatability of retinal oximetry measurements

Skoða venjulega færslu

dc.contributor Landspitali - The National University Hospital of Iceland
dc.contributor.author Karlsson, Róbert Arnar
dc.contributor.author Ólafsdottir, Ólöf Birna
dc.contributor.author Helgadóttir, Védis
dc.contributor.author Belhadj, Soumaya
dc.contributor.author Elíasdóttir, Þórunn Scheving
dc.contributor.author Stefánsson, Einar
dc.contributor.author Harðarson, Sveinn Hákon
dc.date.accessioned 2022-03-19T01:03:41Z
dc.date.available 2022-03-19T01:03:41Z
dc.date.issued 2021-12
dc.identifier.citation Karlsson , R A , Ólafsdottir , Ó B , Helgadóttir , V , Belhadj , S , Elíasdóttir , Þ S , Stefánsson , E & Harðarson , S H 2021 , ' Automation improves repeatability of retinal oximetry measurements ' , PLoS ONE , vol. 16 , no. 12 December , e0260120 . https://doi.org/10.1371/journal.pone.0260120
dc.identifier.issn 1932-6203
dc.identifier.other 44988711
dc.identifier.other 1c884890-9400-47fd-bce9-6b4e467db801
dc.identifier.other 85122015079
dc.identifier.uri https://hdl.handle.net/20.500.11815/2962
dc.description University of Iceland Research Fund (RAK), Icelandic Research Fund (VH), Marie Skłodowska-Curie Actions (SB). Other authors received no funding related to this study (OBO, TSE, SHH). The funders provided support in the form of salaries for authors RAK, VH and SB, but did, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the “author contributions” section. Publisher Copyright: Copyright: © 2021 Karlsson et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
dc.description.abstract Purpose Retinal oximetry is a technique based on spectrophotometry where images are analyzed with software capable of calculating vessel oxygen saturation and vessel diameter. In this study, the effect of automation of measurements of retinal vessel oxygen saturation and vessel diameter is explored. Methods Until now, operators have had to choose each vessel segment to be measured explicitly. A new, automatic version of the software automatically selects the vessels once the operator defines a measurement area. Five operators analyzed image pairs from the right eye of 23 healthy subjects with semiautomated retinal oximetry analysis software, Oxymap Analyzer (v2.5.1), and an automated version (v3.0). Inter- and intra-operator variability was investigated using the intraclass correlation coefficient (ICC) between oxygen saturation measurements of vessel segments in the same area of the retina. Results For semiautomated saturation measurements, the inter-rater ICC was 0.80 for arterioles and venules. For automated saturation measurements, the inter-rater ICC was 0.97 for arterioles and 0.96 for venules. For semiautomated diameter measurements, the inter-rater ICC was 0.71 for arterioles and venules. For automated diameter measurements the inter-rater ICC was 0.97 for arterioles and 0.95 for venules. The inter-rater ICCs were different (p < 0.01) between the semiautomated and automated version in all instances. Conclusion Automated measurements of retinal oximetry values are more repeatable compared to measurements where vessels are selected manually.
dc.format.extent 1664653
dc.format.extent
dc.language.iso en
dc.relation.ispartofseries PLoS ONE; 16(12 December)
dc.rights info:eu-repo/semantics/openAccess
dc.subject Sjónhimna
dc.subject Súrefnismælingar
dc.subject Súrefnismælingar
dc.subject Augnsjúkdómar
dc.subject Náttúrufræðingar
dc.subject Skurð-, svæfinga- og gjörgæsluhjúkrun
dc.subject Augnlæknisfræði
dc.subject Oximetry / methods
dc.subject Oxygen / analysis
dc.subject Retinal Vessels / chemistry
dc.subject Software
dc.subject Multidisciplinary
dc.title Automation improves repeatability of retinal oximetry measurements
dc.type /dk/atira/pure/researchoutput/researchoutputtypes/contributiontojournal/article
dc.description.version Peer reviewed
dc.identifier.doi 10.1371/journal.pone.0260120
dc.relation.url http://www.scopus.com/inward/record.url?scp=85122015079&partnerID=8YFLogxK
dc.contributor.department Faculty of Medicine
dc.contributor.department Other departments
dc.contributor.department Perioperative Services
dc.contributor.department Faculty of Nursing and Midwifery


Skrár

Þetta verk birtist í eftirfarandi safni/söfnum:

Skoða venjulega færslu