Automation improves repeatability of retinal oximetry measurements

Automation improves repeatability of retinal oximetry measurements


Titill: Automation improves repeatability of retinal oximetry measurements
Höfundur: Karlsson, Róbert Arnar
Ólafsdottir, Ólöf Birna
Helgadóttir, Védis
Belhadj, Soumaya
Elíasdóttir, Þórunn Scheving
Stefánsson, Einar
Harðarson, Sveinn Hákon
Útgáfa: 2021-12
Tungumál: Enska
Umfang: 1664653
Háskóli/Stofnun: Landspitali - The National University Hospital of Iceland
Deild: Faculty of Medicine
Other departments
Perioperative Services
Faculty of Nursing and Midwifery
Birtist í: PLoS ONE; 16(12 December)
ISSN: 1932-6203
DOI: 10.1371/journal.pone.0260120
Efnisorð: Sjónhimna; Súrefnismælingar; Súrefnismælingar; Augnsjúkdómar; Náttúrufræðingar; Skurð-, svæfinga- og gjörgæsluhjúkrun; Augnlæknisfræði; Oximetry / methods; Oxygen / analysis; Retinal Vessels / chemistry; Software; Multidisciplinary
URI: https://hdl.handle.net/20.500.11815/2962

Skoða fulla færslu

Tilvitnun:

Karlsson , R A , Ólafsdottir , Ó B , Helgadóttir , V , Belhadj , S , Elíasdóttir , Þ S , Stefánsson , E & Harðarson , S H 2021 , ' Automation improves repeatability of retinal oximetry measurements ' , PLoS ONE , vol. 16 , no. 12 December , e0260120 . https://doi.org/10.1371/journal.pone.0260120

Útdráttur:

Purpose Retinal oximetry is a technique based on spectrophotometry where images are analyzed with software capable of calculating vessel oxygen saturation and vessel diameter. In this study, the effect of automation of measurements of retinal vessel oxygen saturation and vessel diameter is explored. Methods Until now, operators have had to choose each vessel segment to be measured explicitly. A new, automatic version of the software automatically selects the vessels once the operator defines a measurement area. Five operators analyzed image pairs from the right eye of 23 healthy subjects with semiautomated retinal oximetry analysis software, Oxymap Analyzer (v2.5.1), and an automated version (v3.0). Inter- and intra-operator variability was investigated using the intraclass correlation coefficient (ICC) between oxygen saturation measurements of vessel segments in the same area of the retina. Results For semiautomated saturation measurements, the inter-rater ICC was 0.80 for arterioles and venules. For automated saturation measurements, the inter-rater ICC was 0.97 for arterioles and 0.96 for venules. For semiautomated diameter measurements, the inter-rater ICC was 0.71 for arterioles and venules. For automated diameter measurements the inter-rater ICC was 0.97 for arterioles and 0.95 for venules. The inter-rater ICCs were different (p < 0.01) between the semiautomated and automated version in all instances. Conclusion Automated measurements of retinal oximetry values are more repeatable compared to measurements where vessels are selected manually.

Athugasemdir:

University of Iceland Research Fund (RAK), Icelandic Research Fund (VH), Marie Skłodowska-Curie Actions (SB). Other authors received no funding related to this study (OBO, TSE, SHH). The funders provided support in the form of salaries for authors RAK, VH and SB, but did, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the “author contributions” section. Publisher Copyright: Copyright: © 2021 Karlsson et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.