The design of RIP belts impacts the reliability and quality of the measured respiratory signals

dc.contributor.authorMontazeri, Kristofer
dc.contributor.authorJonsson, Sigurdur Aegir
dc.contributor.authorAgustsson, Jon Skirnir
dc.contributor.authorSerwatko, Marta
dc.contributor.authorGislason, Thorarinn
dc.contributor.authorArnardottir, Erna S.
dc.contributor.departmentDepartment of Engineering
dc.contributor.departmentDepartment of Computer Science
dc.date.accessioned2026-11-11T11:43:01Z
dc.date.available2026-11-11T11:43:01Z
dc.date.issued2021-09
dc.descriptionPublisher Copyright: © 2021, The Author(s).en
dc.description.abstractPurpose: Evaluate the effect of respiratory inductance plethysmography (RIP) belt design on the reliability and quality of respiratory signals. A comparison of cannula flow to disposable cut-to-fit, semi-disposable folding and disposable RIP belts was performed in clinical home sleep apnea testing (HSAT) studies. Methods: This was a retrospective study using clinical HSAT studies. The signal reliability of cannula, thorax, and abdomen RIP belts was determined by automatically identifying periods during which the signals did not represent respiratory airflow and breathing movements. Results were verified by manual scoring. RIP flow quality was determined by examining the correlation between the RIP flow and cannula flow when both signals were considered reliable. Results: Of 767 clinical HSAT studies, mean signal reliability of the cut-to-fit, semi-disposable, and disposable thorax RIP belts was 83.0 ± 26.2%, 76.1 ± 24.4%, and 98.5 ± 9.3%, respectively. The signal reliability of the cannula was 92.5 ± 16.1%, 87.0 ± 23.3%, and 85.5 ± 24.5%, respectively. The automatic assessment of signal reliability for the RIP belts and cannula flow had a sensitivity of 50% and a specificity of 99% compared with manual assessment. The mean correlation of cannula flow to RIP flow from the cut-to-fit, semi-disposable, and disposable RIP belts was 0.79 ± 0.24, 0.52 ± 0.20, and 0.86 ± 0.18, respectively. Conclusion: The design of RIP belts affects the reliability and quality of respiratory signals. The disposable RIP belts that had integrated contacts and did not fold on top of themselves performed the best. The cut-to-fit RIP belts were most likely to be unreliable, and the semi-disposable folding belts produced the lowest-quality RIP flow signals compared to the cannula flow signal.en
dc.description.versionPeer revieweden
dc.format.extent7
dc.format.extent417970
dc.format.extent1535-1541
dc.identifier.citationMontazeri, K, Jonsson, S A, Agustsson, J S, Serwatko, M, Gislason, T & Arnardottir, E S 2021, 'The design of RIP belts impacts the reliability and quality of the measured respiratory signals', Sleep and Breathing, vol. 25, no. 3, pp. 1535-1541. https://doi.org/10.1007/s11325-020-02268-xen
dc.identifier.doi10.1007/s11325-020-02268-x
dc.identifier.issn1520-9512
dc.identifier.other251576821
dc.identifier.other5adcc35c-b57b-45ef-8a98-43f1bffc27eb
dc.identifier.other85099109612
dc.identifier.other33411184
dc.identifier.urihttps://hdl.handle.net/20.500.11815/9072
dc.language.isoen
dc.relation.ispartofseriesSleep and Breathing; 25(3)en
dc.relation.urlhttps://www.scopus.com/pages/publications/85099109612en
dc.rightsinfo:eu-repo/semantics/openAccessen
dc.subjectObstructive sleep apnea (OSA)en
dc.subjectRespiratory airflowen
dc.subjectRespiratory inductance plethysmography (RIP) beltsen
dc.subjectSignal qualityen
dc.subjectSignal reliabilityen
dc.subjectOtorhinolaryngologyen
dc.subjectNeurology (clinical)en
dc.titleThe design of RIP belts impacts the reliability and quality of the measured respiratory signalsen
dc.type/dk/atira/pure/researchoutput/researchoutputtypes/contributiontojournal/articleen

Skrár

Original bundle

Niðurstöður 1 - 1 af 1
Nafn:
s11325-020-02268-x.pdf
Stærð:
408.17 KB
Snið:
Adobe Portable Document Format