Opin vísindi

Self-applied somnography : technical feasibility of electroencephalography and electro-oculography signal characteristics in sleep staging of suspected sleep-disordered adults

Skoða venjulega færslu

dc.contributor Landspitali - The National University Hospital of Iceland
dc.contributor.author Rusanen, Matias
dc.contributor.author Korkalainen, Henri
dc.contributor.author Grétarsdóttir, Heiður
dc.contributor.author Siilak, Tiina
dc.contributor.author Olafsdottir, Kristin Anna
dc.contributor.author Töyräs, Juha
dc.contributor.author Myllymaa, Sami
dc.contributor.author Arnardóttir, Erna Sif
dc.contributor.author Leppänen, Timo
dc.contributor.author Kainulainen, Samu
dc.date.accessioned 2023-12-06T01:04:58Z
dc.date.available 2023-12-06T01:04:58Z
dc.date.issued 2023-07-03
dc.identifier.citation Rusanen , M , Korkalainen , H , Grétarsdóttir , H , Siilak , T , Olafsdottir , K A , Töyräs , J , Myllymaa , S , Arnardóttir , E S , Leppänen , T & Kainulainen , S 2023 , ' Self-applied somnography : technical feasibility of electroencephalography and electro-oculography signal characteristics in sleep staging of suspected sleep-disordered adults ' , Journal of Sleep Research . https://doi.org/10.1111/jsr.13977
dc.identifier.issn 0962-1105
dc.identifier.other 172187205
dc.identifier.other 9b117afe-62d5-4f05-beeb-0d48a3e35cc2
dc.identifier.other 85164208096
dc.identifier.other unpaywall: 10.1111/jsr.13977
dc.identifier.uri https://hdl.handle.net/20.500.11815/4583
dc.description Funding Information: Financial support for this study was provided by the European Union's Horizon 2020 Research and Innovation Programme under grant agreement No 965417, by the Research Committee of the Kuopio University Hospital Catchment Area for the State Research Funding (Grants 5041807, 5041804, 5041803, 5041797, and 5041794), by the Finnish Cultural Foundation through Kainuu Regional Fund and Central fund, by Olvi Foundation, by the Finnish Anti‐Tuberculosis Association, by Tampere Tuberculosis Foundation, by the Research Foundation of the Pulmonary Diseases, by the NordForsk (NordSleep Project 90458) through the Business Finland (Grant 5133/31/2018), by the Kuopio University Hospital Research Foundation, and The Icelandic Centre for Research. Publisher Copyright: © 2023 The Authors. Journal of Sleep Research published by John Wiley & Sons Ltd on behalf of European Sleep Research Society.
dc.description.abstract Sleep recordings are increasingly being conducted in patients’ homes where patients apply the sensors themselves according to instructions. However, certain sensor types such as cup electrodes used in conventional polysomnography are unfeasible for self-application. To overcome this, self-applied forehead montages with electroencephalography and electro-oculography sensors have been developed. We evaluated the technical feasibility of a self-applied electrode set from Nox Medical (Reykjavik, Iceland) through home sleep recordings of healthy and suspected sleep-disordered adults (n = 174) in the context of sleep staging. Subjects slept with a double setup of conventional type II polysomnography sensors and self-applied forehead sensors. We found that the self-applied electroencephalography and electro-oculography electrodes had acceptable impedance levels but were more prone to losing proper skin–electrode contact than the conventional cup electrodes. Moreover, the forehead electroencephalography signals recorded using the self-applied electrodes expressed lower amplitudes (difference 25.3%–43.9%, p < 0.001) and less absolute power (at 1–40 Hz, p < 0.001) than the polysomnography electroencephalography signals in all sleep stages. However, the signals recorded with the self-applied electroencephalography electrodes expressed more relative power (p < 0.001) at very low frequencies (0.3–1.0 Hz) in all sleep stages. The electro-oculography signals recorded with the self-applied electrodes expressed comparable characteristics with standard electro-oculography. In conclusion, the results support the technical feasibility of the self-applied electroencephalography and electro-oculography for sleep staging in home sleep recordings, after adjustment for amplitude differences, especially for scoring Stage N3 sleep.
dc.format.extent 2421657
dc.format.extent
dc.language.iso en
dc.relation.ispartofseries Journal of Sleep Research; ()
dc.rights info:eu-repo/semantics/openAccess
dc.subject Náttúrufræðingar
dc.subject home sleep apnea testing
dc.subject portable diagnostics
dc.subject sleep-disordered breathing
dc.subject Cognitive Neuroscience
dc.subject Behavioral Neuroscience
dc.title Self-applied somnography : technical feasibility of electroencephalography and electro-oculography signal characteristics in sleep staging of suspected sleep-disordered adults
dc.type /dk/atira/pure/researchoutput/researchoutputtypes/contributiontojournal/article
dc.description.version Peer reviewed
dc.identifier.doi 10.1111/jsr.13977
dc.relation.url http://www.scopus.com/inward/record.url?scp=85164208096&partnerID=8YFLogxK
dc.contributor.department Department of Engineering


Skrár

Þetta verk birtist í eftirfarandi safni/söfnum:

Skoða venjulega færslu